Lift-U Wheelchair Lift – 06E093000

Defect Description For Recall

Certain Lift-U Wheelchair Lifts Equipped With A Solid Bar Barrier Linkage Installed As Original Equipment On Certain Specialty Vehicles And Also Sold As Aftermarket Retrofit Equipment. The Part Numbers Which Make Up The Solid Bar Linkage Are: 184-0042, 184-0041, 123-0288, And 123-0287. It Is Possible For A Person To Wedge Their Feet Under The Linkage.

Make Of Vehicle

LIFT-U

Model Of Vehicle

WHEELCHAIR LIFT

Year

9999

Recall Campaign Number

06E093000

Component Description

Equipment Adaptive/Mobility

Effect Of Defect

It Is Possible That A Person With No Shoes, No Control Of Their Foot Could Twist Their Foot And Get It Forced Under The Linkage While Setting In A Wheel Chair Which Could Result In A Cut.

Fix Defect

The Vehicle Manufacturers Will Notify Owners To Contact Lift-U For A Free Repair Kit Along With Installation InstruCTIons. Lift-U Will Notify Owners Who Have Purchased The Subject Lifts As Aftermarket Equipment. The Recall Began On January 15, 2007. Owners May Contact Lift-U At 209-838-2400.

Vehicle Manufacturer

Lift-U

Start Of Manufacture

End Of Manufacture

Type Of Vehicle Recall

E

Number Affected

18093

Date Owners Notified

20070117

Initiated By

Mfr

Responsible Manufacturer

Lift-U

Date Report Received

20061030

Make Campaign Number

Part Number

Federal Motor Vehicle Safety Standard